ResearchSpace

Direct activity determination of Mn-54 and Zn-65 by a non-extrapolation liquid scintillation method

Show simple item record

dc.contributor.author Simpson, BRS en_US
dc.contributor.author Morris, WM en_US
dc.date.accessioned 2007-02-08T06:54:27Z en_US
dc.date.accessioned 2007-06-07T10:08:53Z
dc.date.available 2007-02-08T06:54:27Z en_US
dc.date.available 2007-06-07T10:08:53Z
dc.date.copyright en_US
dc.date.issued 2004-02 en_US
dc.identifier.citation Simpson, BRS and Morris, WM. 2004. Direct activity determination of Mn-54 and Zn-65 by a non-extrapolation liquid scintillation method. Applied Radiation and Isotopes, vol. 60, 04 February, pp 475-479 en_US
dc.identifier.issn 0969-8043 en_US
dc.identifier.uri http://hdl.handle.net/10204/1629 en_US
dc.identifier.uri http://hdl.handle.net/10204/1629
dc.description.abstract The measurement of Mn-54 and Zn-65 by liquid scintillation coincidence counting results in low detection efficiencies. The activity obtained from the extrapolation of efficiency data can therefore become problematic if curvature is present. The simple decay scheme exhibited by these radionuclides, with the emission of an energetic gamma ray, allows the absolute activity to be determined from 4pie-gamma data by direct calculation without the need for efficiency extrapolation. The method, which relies on determining the probability of the gamma-ray interacting with the scintillator solution, is described and validated by measurements made on Co-60 en_US
dc.format.extent 194147 bytes en_US
dc.format.mimetype application/pdf en_US
dc.language.iso en en_US
dc.publisher Pergamon-Elsevier Science Ltd en_US
dc.rights Copyright: 2004 Pergamon-Elsevier Science Ltd en_US
dc.source en_US
dc.subject Liquid scintillation coincidence counting en_US
dc.subject Gamma-ray interaction probability en_US
dc.subject Chemistry en_US
dc.subject Mn-54 en_US
dc.subject Zn-65 en_US
dc.title Direct activity determination of Mn-54 and Zn-65 by a non-extrapolation liquid scintillation method en_US
dc.type Article en_US
dc.identifier.apacitation Simpson, B., & Morris, W. (2004). Direct activity determination of Mn-54 and Zn-65 by a non-extrapolation liquid scintillation method. http://hdl.handle.net/10204/1629 en_ZA
dc.identifier.chicagocitation Simpson, BRS, and WM Morris "Direct activity determination of Mn-54 and Zn-65 by a non-extrapolation liquid scintillation method." (2004) http://hdl.handle.net/10204/1629 en_ZA
dc.identifier.vancouvercitation Simpson B, Morris W. Direct activity determination of Mn-54 and Zn-65 by a non-extrapolation liquid scintillation method. 2004; http://hdl.handle.net/10204/1629. en_ZA
dc.identifier.ris TY - Article AU - Simpson, BRS AU - Morris, WM AB - The measurement of Mn-54 and Zn-65 by liquid scintillation coincidence counting results in low detection efficiencies. The activity obtained from the extrapolation of efficiency data can therefore become problematic if curvature is present. The simple decay scheme exhibited by these radionuclides, with the emission of an energetic gamma ray, allows the absolute activity to be determined from 4pie-gamma data by direct calculation without the need for efficiency extrapolation. The method, which relies on determining the probability of the gamma-ray interacting with the scintillator solution, is described and validated by measurements made on Co-60 DA - 2004-02 DB - ResearchSpace DP - CSIR KW - Liquid scintillation coincidence counting KW - Gamma-ray interaction probability KW - Chemistry KW - Mn-54 KW - Zn-65 LK - https://researchspace.csir.co.za PY - 2004 SM - 0969-8043 T1 - Direct activity determination of Mn-54 and Zn-65 by a non-extrapolation liquid scintillation method TI - Direct activity determination of Mn-54 and Zn-65 by a non-extrapolation liquid scintillation method UR - http://hdl.handle.net/10204/1629 ER - en_ZA