De Vreede, JPMKorfage, WPersson, PBrunetti, LLopez, VPetras, IMorard, PHejsek, FTorok, ARuhaak, JAscroft, JDressler, ECelep, MLapuh, RAchkar, J2007-02-082007-06-072007-02-082007-06-072001-04De Vreede, JPM, et al. 2001. International comparison for RF power in the frequency range up to 18 GHz. IEEE Transactions on Instrumentation and Measurement, vol. 50(2), pp 409-413.0018-9456http://hdl.handle.net/10204/1671http://hdl.handle.net/10204/1671A Euromet international laboratory comparison (Project 393) has been carried out between 14 national standards laboratories. Thermistor mounts were used, equipped with PC7 as well as Type N connectors. The comparison is carried out using the normal equipment of the laboratory for high-quality external calibration. The results show good agreement in measuring the calibration coefficient of the thermistor mounts within the claimed expanded uncertainty (typically between 1% and 2%). It confirms the equivalence of national standards for RF power up to 18 GHz. In one case, corrective action is proposed.102090 bytesapplication/pdfenCopyright: 2001 Institute of Electrical and Electronics Engineers Inc.EngineeringEquivalence of standardsMeasurement standardsPower measurementsThermistorsTransfer standardsInternational comparison for RF power in the frequency range up to 18 GHzArticleDe Vreede, J., Korfage, W., Persson, P., Brunetti, L., Lopez, V., Petras, I., ... Achkar, J. (2001). International comparison for RF power in the frequency range up to 18 GHz. http://hdl.handle.net/10204/1671De Vreede, JPM, W Korfage, P Persson, L Brunetti, V Lopez, I Petras, P Morard, et al "International comparison for RF power in the frequency range up to 18 GHz." (2001) http://hdl.handle.net/10204/1671De Vreede J, Korfage W, Persson P, Brunetti L, Lopez V, Petras I, et al. International comparison for RF power in the frequency range up to 18 GHz. 2001; http://hdl.handle.net/10204/1671.TY - Article AU - De Vreede, JPM AU - Korfage, W AU - Persson, P AU - Brunetti, L AU - Lopez, V AU - Petras, I AU - Morard, P AU - Hejsek, F AU - Torok, A AU - Ruhaak, J AU - Ascroft, J AU - Dressler, E AU - Celep, M AU - Lapuh, R AU - Achkar, J AB - A Euromet international laboratory comparison (Project 393) has been carried out between 14 national standards laboratories. Thermistor mounts were used, equipped with PC7 as well as Type N connectors. The comparison is carried out using the normal equipment of the laboratory for high-quality external calibration. The results show good agreement in measuring the calibration coefficient of the thermistor mounts within the claimed expanded uncertainty (typically between 1% and 2%). It confirms the equivalence of national standards for RF power up to 18 GHz. In one case, corrective action is proposed. DA - 2001-04 DB - ResearchSpace DP - CSIR KW - Engineering KW - Equivalence of standards KW - Measurement standards KW - Power measurements KW - Thermistors KW - Transfer standards LK - https://researchspace.csir.co.za PY - 2001 SM - 0018-9456 T1 - International comparison for RF power in the frequency range up to 18 GHz TI - International comparison for RF power in the frequency range up to 18 GHz UR - http://hdl.handle.net/10204/1671 ER -