A Euromet international laboratory comparison (Project 393) has been carried out between 14 national standards laboratories. Thermistor mounts were used, equipped with PC7 as well as Type N connectors. The comparison is carried out using the normal equipment of the laboratory for high-quality external calibration. The results show good agreement in measuring the calibration coefficient of the thermistor mounts within the claimed expanded uncertainty (typically between 1% and 2%). It confirms the equivalence of national standards for RF power up to 18 GHz. In one case, corrective action is proposed.
Reference:
De Vreede, JPM, et al. 2001. International comparison for RF power in the frequency range up to 18 GHz. IEEE Transactions on Instrumentation and Measurement, vol. 50(2), pp 409-413.
De Vreede, J., Korfage, W., Persson, P., Brunetti, L., Lopez, V., Petras, I., ... Achkar, J. (2001). International comparison for RF power in the frequency range up to 18 GHz. http://hdl.handle.net/10204/1671
De Vreede, JPM, W Korfage, P Persson, L Brunetti, V Lopez, I Petras, P Morard, et al "International comparison for RF power in the frequency range up to 18 GHz." (2001) http://hdl.handle.net/10204/1671
De Vreede J, Korfage W, Persson P, Brunetti L, Lopez V, Petras I, et al. International comparison for RF power in the frequency range up to 18 GHz. 2001; http://hdl.handle.net/10204/1671.